Number of Pixels: 2584 (H) x 1936 (V) = 5 Megapixel Digitization: 12-bit / 12 Mhz pixel clock Dynamic Ranbge: 1:1300 Integration time: 1 ms to 60 s Cooling: Single stage Peltier cooling Interface: FireWire / IEEE 1394, 6 pin jack, speed 400 Mbit
Live Image Frame Rate:
Mode
H
x
V
Max.Frame Rate
Slow
1296
x
968
3 frames/s
Medium
430
x
322
10 frames/s
Fast
258
x
193
14 frames/s
Note: Max. image frame rate dependes on the exposure time and the hard- and software of the PC.
Selectable Resolution in Quality Mode (Interlaced): (Recommended for highest image quality)
H
x
V
Max. Frame Rate
516
x
387
5 x 5
646
x
484
4 x 4
861
x
645
3 x 3
1292
x
968
2 x 2
2584
x
1936
1 x 1
(all Modes in color)
Selectable Resolution in Fast Mode (Progressive): (Recommended for moving objects)
H
x
V
Binning
Farbe
258
x
193
10 x 10
Color
323
x
242
8 x 8
Mono
430
x
322
6 x 6
Color
646
x
484
4 x 4
Mono
1292
x
968
2 x 2
Color
1292
x
968
2 x 2
Mono
2584
x
1936
1 x 1, interpolate d
Color
Read-out of sensor sub-regions: Random definition of regions of interest (ROI) on the sensor.
Pixel size
3,4 µm x 3,4 µm
Sensor size
8,7 mm x 6,6 mm equivalent to 2/3" CCD
Sensor
ICX 282, Interline Transfer
Spectral Range
Approx. 400 nm to 710 nm with infrared barrier filter BG40
Read-out modes
Progressive / Interlaced
Control signals
TTL-output for controlling external electrical shutters
Optical interface
C-mount
Thread depth for objectives
Max. of 5 mm (C-mount)
Max. file size per image
Approx. 30 MB at 2584 x 1936 with 3 x 12 bit
Operating systems
Windows 2000 Professional (from SP3) Windows XP Professional (from SP1)
Size / weight
Approx. 11 cm x 8 cm x 4,5 cm / 370 g
Housing
Blue anodized aluminum, with cooling fins, 1/4" connection for tripod mount
Registration
CE, cUL
Power Supply
12 V DC, 250 mA, supply via FireWire / IEEE 1394 interface and data cable from the PC (no external power supply necessary if used in combination with PC). If camera is used together with a notebook a FireWire / IEEE 1394 hub is required.
Ambient condition (operation)
+5° to +35° Celsius; max. of 80% relative air humidity, not condensing Free air circulation required